Agilent Technologies 1100 Series Reference Manual page 152

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector
5
Insert the new or cleaned condenser, shift it completely
to its mechanical stop and fit the screw.
Next Steps:
Replace the covers as described in "Replacing the Foam and the Top Cover" on page 149.
Re-install the detector in the stack.
Enter the replacement of the lamp in the maintenance logbook (this will reset the lamp counter).
Turn the lamp on.
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
"Wavelength Verification and Calibration"
154
6
Replace the lamp housing cover and fix the screws.
Carefully insert the rear foam. Reconnect the
monochromator cables to the main board. Left is
excitation, right is emission.
Rear
foam
EM
EX
on page 101.
A
A
A
B
1100 Series FD Reference Manual