Performance Specifications - Agilent Technologies 1100 Series User Manual

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

1
Site Requirements and Specifications

Performance Specifications

Table 2
Performance Specifications Agilent 1100 Series DAD and MWD
Type
Detection type
Light source
Wavelength range
Short term noise (ASTM)
Single and
Multi-Wavelength
Drift
Linear absorbance range
Wavelength accuracy
Wavelength bunching
Slit width
Diode width
Flow cells
Control and data evaluation
16
Specification
1024-element photodiode array
Deuterium and tungsten lamps
190 – 950 nm
-5
± 1 × 10
AU at 254 and 750 nm
-3
2 × 10
AU/hr at 254 nm
> 2 AU (upper limit)
± 1 nm
1 – 400 nm
1, 2, 4 , 8, 16 nm
< 1 nm
Standard: 13 µl volume, 10 mm cell path length
and 120 bar (1760 psi) pressure maximum
Semi-Micro: 5 µl volume, 6 mm cell path length
and 120 bar (1760 psi) pressure maximum
High pressure: 1.7 µl volume, 6 mm cell path
length and 400 bar (5880 psi) pressure maximum
500 nano: 0.5 µl volume, 10 mm cell path length
and 50 bar (725 psi) pressure maximum
80 nano: 0.5 µl volume, 10 mm cell path length and
50 bar (725 psi) pressure maximum
Agilent ChemStation for LC (32-bit)
Comments
The UV-lamp is equipped with I.D. tag
that holds lamp typical information.
See note on page 17
See note on page 17
See note on page 17
Self-calibration with deuterium lines,
verification with holmium oxide filter
Programmable in steps of 1 nm
Programmable slit
See
"Optimization Overview"
page 89
All flow cells are equipped with I.D.
tags that hold cell typical information.
Revision B.01.03 or above
1100 Series DAD and MWD User Manual
on