Performance Specifications Agilent 1100 Series Fluorescence Detector
Specification
Excitation or Emission spectra
Scan speed: 28 ms per datapoint
(e.g. 0.6 s/spectrum 200-400 nm, 10 nm step)
Step size: 1-20 nm
Spectra storage: All
Repeatability+/- 0.2 nm
Accuracy+/- 3 nm setting
Standard: 8 µl volume and 20 bar
(2 MPa) pressure maximum,
quartz
Optional:
Fluorescence cuvette for offline
spectroscopic measurements
with 1 ml syringe, 8 µl volume,
quartz
Agilent ChemStation for LC,
Agilent 1100 Control Module with
limited spectral data analysis and
printing of spectra
Recorder/integrator: 100 mV or
1 V, output range >10
luminescence units, two outputs
Controller-area network (CAN),
GPIB, RS-232C, LAN, APG
Remote: ready, start, stop and
shut-down signals
Extensive diagnostics, error
detection and display (through
control module and ChemStation),
leak detection, safe leak handling,
leak output signal for shutdown of
pumping system. Low voltages in
major maintenance areas.
Comments
2
10
Specifications
329