Wavelength Calibration Failed
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message "Wavelength Calibration Failed" is displayed and the instrument
stays in a Not Ready condition.
Probable Causes
• Flash lamp not ignited or position not correct.
• Cell position not correct.
• Solvent in the cell not clean or air bubble in the cell.
• monochromator assembly position not correct (after replacement).
Suggested Actions
1100 Series FD Reference Manual
Check the flash lamp image and position. See also
Check the cell position.
Flush the flow cell.
Reset monochromator settings and re-run the wavelength calibration.
Troubleshooting and Test Functions
"No Peaks"
4
on page 100.
97