10
Replace the front foam.
1100 Series FD Reference Manual
Repairing the Fluorescence Detector
Next Steps:
Replace the covers as described in
Foams and Covers"
on page 196.
Re-install the flow cell and the front panel
Re-install the detector in the stack.
Turn the lamp on.
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
"Wavelength Verification and Calibration"
page 101.