4
Troubleshooting and Test Functions
9 You can now select a stored spectra and evaluate/print them.
10 Load the method WLEMTEST. The FLD will change into the multi-excitation
11 Repeat steps
Figure 39
12 You can now select a stored spectra and evaluate/print them.
The spectra on the control module is a real raw data spectrum. Therefore it looks not that
NO TE
smooth as on the Agilent ChemStation (uses the spline algorithm).
108
The maximum should be found at 397 nm
mode and scan in the range of the expected maximum of 350 nm
step 6
Evaluating Online Excitation Spectra
The maximum should be found at 350 nm
to
step
8.
±
3 nm.
±
3 nm.
1100 Series FD Reference Manual
±
20 nm.