Replacing FLL board and Trigger Pack

When required
Tools required
Parts required
When working on the optical unit, a clean workbench with ESD protection mat must be
CAU TI O N
available. Otherwise optical components or electronic boards may be damaged.
The instrument should not be operated without Xenon flash lamp. Otherwise the
WA RN ING
Triggerpack electronics will be damaged.
Preparations for this procedure:
Optical unit has been removed as described in
"Removing the Optical Unit"
The foam must be in place. Otherwise cutoff filter and
diffuser will fall out.
1100 Series FD Reference Manual
If defective
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
FLL Board G1321-66512 (includes trigger pack)
on page 178.
Repairing the Fluorescence Detector
1 Place the optical unit on the ESD protection mat.