Agilent Technologies 1100 Series Reference Manual page 121

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Set the
Stoptime = 23
minutes, see
on
page
121.
Figure 51
Interpretation of the Results
If the test shows low Raman values, check for:
1100 Series FD Reference Manual
Raman Test - FLD Settings (part of method)
correctly positioned flow cell,
clean flow cell (flush),
no air bubble(s) (check via fluorescence scan or visual check of
cell/cuvette),
solvent inlet filter (may create air bubbles in flow cell),
clean water.
Troubleshooting and Test Functions
4
123