Introduction to the Fluorescence Detector
length of the boxcar element using the RESPONSETIME function: the longer
the RESPONSETIME, the greater the number of data points averaged. A
four-fold increase in RESPONSETIME (for example, 1 sec to 4 sec) doubles the
signal-to-noise ratio.
Figure 87
RESPONSETIME: Signal-to-Noise Ratio
1100 Series FD Reference Manual