Exchanging the Fan
When required
Tools required
Parts required
The fan must be installed in the correct orientation to ensure optimum cooling and
CAU TI O N
operation of the detector.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
Covers"
on page 144.
1100 Series FD Reference Manual
If the fan is defective or noisy
Screwdriver POZI 1 PT3
Fan assembly 3160-1017
1
"Removing the
EM
EX
Repairing the Fluorescence Detector
Disconnect the emission and the excitation cables and
from the detector main board.
A
A
A
B
5
B
163