Removing The Covers - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector

Removing the Covers

When required
Tools required
Parts required
There will be a risk of damaging hardware due to overheating when operating the
CAU TI O N
instrument without covers.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
144
For all repairs inside the detector
Screwdriver POZI 1 PT3
Depends on the work inside and the following procedures
1
Press the release buttons and remove the front cover.
Unclip and remove the waste funnel assemly.
1100 Series FD Reference Manual