Exchanging the multi-channel gradient valve on the quaternary pump (4 pages)
Summary of Contents for Agilent Technologies 1100 series
Page 1
System Quick Start Guide A guide to help you get started with the LC/MSD Trap system What is the Agilent 1100 Series LC/MSD Trap System? Where to find additional system information What is new in version 5.2 software? Instructional overview Step 1—Start the software...
What is the Agilent 1100 Series LC/MSD Trap System? The LC/MSD Trap system consists of the Agilent 1100 Series LC, an atmospheric pressure ion source and the MSD Trap mass spectrometer, which can be one of several models: Classic, VL, SL, or XCT.
Presents instructions to back up and recover the PC, including OS and application software, if there is a problem. Included with new system bundles; also available on the CD as a PDF file. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
New for individual programs with or without compliance software LC/CE/Sequence control with ChemStation A.10.02 • Capacity to store methods in any location (recommended to use subdirectories in the default method root) Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
• Improved Find Compounds - AutoMS(n) algorithm (limits algorithm to find the x most intense compounds) • Improved monoisotopic peak determination of precursor • Library Search: Inclusion of the measured mass range with the spectrum properties. QuantAnalysis • Improved weighted calibration Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Instructional overview Install the Trap System hardware Use the instructions in the Agilent 1100 Series LC/MSD Trap System Installation Guide to install the hardware. Install the Trap System software Use the instructions in the Agilent 1100 Series LC/MSD Trap System Installation Guide to install all the software components.
The instructions below include the following assumptions: • The hardware and software are properly installed. • The instrument is configured. Use instructions in the Agilent 1100 Series LC/MSD Trap System Installation Guide to configure the instrument for the first time.
Page 8
MSD Trap Control. ChemStation MSD Trap Control • Click the ChemStation button in MSD Trap Control. DataAnalysis • Select Tools > ChemStation, or • Click the ChemStation button in DataAnalysis. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
The ReportDesigner program lets you create your own report layouts or modify the existing ones. • Select Start > Programs > LCMSD Trap > ReportDesigner, 1 Open DataAnalysis. 2 Select Print..., or Print Preview..., > ReportDesigner..Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
LC part of LC-only or LC/MS methods, set up information for single runs or sequences and acquire data. See the Agilent ChemStation document set and ChemStation Online Help to familiarize yourself with this software. Figure 1 ChemStation window Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Top Menu Toolbar Line spectra displayed here Profile spectra displayed here Display Panes Chromatograms displayed here Status Modes Tabs General Trap Parameters Acquisition Method Editor Figure 2 Initial MSD Trap Control window Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 12
ChemStation, then the file name and directory locations are prescribed in ChemStation’s Sample Info. Stop the acquisition Acquisition > Stop Run (F8) Pause the acquisition Acquisition > Pause (Ctrl + F8) Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 13
Tile the spectrum windows View > Tile Spectrum Windows Show the LC program View > LC Program (Shift+F11) Show DataAnalysis Program View > DataAnalysis (F11) About MSD Trap Control Help > About... Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 14
To adjust the frame border move the cursor to the edge and click and drag the border until your screen appears as in Figure Figure 3 Comparison of a Line spectrum and Profile spectrum Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 15
See “Exercise 3.Change and run method to optimize peak height” on page 45 of the Familiarization Guide to learn how to set up segments. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 16
This results in a slight degradation of mass assignment and resolution. Suggested values for the ICC Target are provided with the default methods (e.g., DEF_LCMS.M). Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 17
This action improves the spectral quality at the expense of chromatographic fidelity. Peaks appear skewed (rapid rise followed by a longer tailing edge). Disable this feature for quantitative work. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Trap trap. The Divert Valve is switched To Waste. before you are able to switch the ion sources.] Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
7 to learn how to open this window. “Step 4—Review data and identify compounds” on page 28 and “Deconvolute and search protein spectra” on page 34 for more information on how to to use this window. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Figure 6 QuantAnalysis window “Step 1—Start the software” on page 7 to learn how to open this window. “Quantify compounds” on page 35 for more information on how to use this window. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Do a scan Prepare the Start calibration LC to run the here! sample Do an Prepare the isolation calibration instrument Optimize Optimize Do a fragmentation sensitivity fragmentation (MS/MS) (MS) calibration Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Start Enter LC Enter MS parameters parameters Here! Set up and run an Enter sample acquisition information method Acquire data Set up displays Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
If infusing into LC flows, set LC rate to be used during the analysis. pump rate in ChemStation. Figure 7 Tubing from the Syringe Pump (1) to the Nebulizer (2) Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 24
Dry Temp 325 – 350 °C 5 Switch to Operate state. • On the Main window, select the • All of the electrical elements for Operate button. generating and directing ions are turned on. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 25
• You cannot change the name of the b Replace the name of the default default method. You must save all method with the name of your new changes under a new method name. method. c Click OK. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
• The previously selected mass is set window to open a context menu, as the isolation mass. and select Isolate/Fragment. • Check that the ion is isolated and all other ions in the Line Spectrum window disappear. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Sample Info tab. 2 Stop the acquisition. Stop the method manually: • Click the Stop button, or • From the top menu, select RunControl > Stop Run. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Review data for automatic compounds and identify processing and based on compounds reporting spectra Create and (Optional) Generate use library to Create your reports from identify own report default layouts compounds layouts Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Select the trace with the mouse cursor. c Make the desired changes. d Click Change to confirm the changes to the trace. e Click OK. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Line and profile spectra. MS chromatogram. f Select Find > Compounds – Chromatogram. View three spectra in the same a Right-click the Compound Mass window. Spectra window. b Select List windows > 3 windows. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Report. b For this example, select Cmpd Mass Spec Report –AutoMS(n) (P) from the list of Layouts. c Click OK. d When prompted to save your processed results, always click Yes. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Select Identify > Parameters... c Click the Libraries tab. d Click the New (Insert) button in the Libraries tab. e Click the Browse button (...). f Select the database, and click Open. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 33
Print the results from the Compound a In DataAnalysis, select File > List window to a Library Search Report. Print..b Select Library Search Report – AutoMS(n) from the pull-down list. c Click OK. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
• Online Help for the tasks listed for the roadmap below and descriptions of the fields found on the windows and dialog boxes needed for these tasks. Start Deconvolute Extract spectra spectra for Here! deconvolution Deconvolute and search Export spectra protein spectra Search MASCOT database Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Start Create a Import and quantitation modify a Here! method sequence Quantify an Quantify unknown compounds sample Print the quantitation report Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
3 Select Ask which mode to activate, and click Apply. 4 Exit the MSD Trap Control software. 5 When the message appears to ask for Standby or Shutdown, select Standby, and click OK. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
• Power failure to the PC or to the Trap (Upon power restoration, the instrument remains shut down.) • Gas flow error Shutdown mode indicator becomes red. Active run or sequence is stopped. Divert valve is set To Waste. • Vacuum fault Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Demonstrates all properties of UV chromatograms Add_UV.d Add_Var.M Demonstrates all properties of acquisition parameter traces Add_VAR.d Add_Mult.M Creates an array of chromatograms Add_Mult.d Add_Range.M Demonstrates the 'AddRangeSelection' and RangeAvg.d 'AverageMassSpectrum' methods Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 39
Opens another data file Add_BPC.d Base.d RngeSelDiss.M Runs the Find Compounds Dissect function for a specified time S_MS.d range RngeSelAutoMSn.M Runs the Find Compounds AutoMSn function for a specified S_AMS2N1.d time range Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 40
Compound Mass Spectra entry with the averaged spectrum. BaseLineSubtract Add_Range Performs a baseline subtraction on spectra of a Compound Mass Spectra object. The parameters for this algorithm will be taken from method. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 41
Also export spectra to MGF, ASCII, XML, or CSV ExpSpec; Form; Metabolite_ID; Smooth formats and Compound MS to MGF, XML, and CSV formats. ExportMassList ExpMList Exports Compound MS mass lists to ASCII or CSV formats. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 42
Chromatogram. Data Mining of MS and UV data files. Finds peaks within the chromatograms traces and integrates them. Places results in Compound List Open Open_File Opens *.d file from user-specified directory. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 43
Help > VBScript on the Web > Install Microsoft ActiveX Control Pad… This enables the use of a special toolbox for use in the Object View of the VBScript window (see figure below). Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
SL_XCT Options Options and Advanced dialogs VL OneSegmentMRM(2)_2Ions Defines a Method with only one Segment, MRM operations and two iterations of two precursor ions Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
(overlaid), compound mass spectra and corresponding mass lists Cmpd Mass Spec Report - AutoMSn Analysis Info, most important acquisition parameters (P).layout (including AutoMS(n)), chromatogram display (overlaid) and compound mass spectra Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 46
Modified Cmpd Chrom Report 3R1C - MS (P).layout to represent S/N results for installation checkout Single Mass Spec Decon Report Analysis info, important acquisition parameters, (P).layout display of Mass Spectrum window and deconvolution results for spectrum Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
Page 47
† For all Display reports, the windows available to open include Chromatogram, Mass Spectrum View, Compound Mass Spectra and Mass Spectrum windows. Agilent 1100 Series LC/MSD Trap System Quick Start Guide...
• Shut down or place the instrument in standby This book also contains a listing of new 5.2 software features and shipped methods and report layouts. Agilent Technologies, Inc. 2004 Printed in USA First Edition, April 2004 *G2440-90089 * G2440-90089 Agilent Technologies...