11 Reinstall the top foam section, top cover and front cover, see
12 Replace detector into the stack and reconnect the cables.
If a new FLM board is installed, update the serial number information of the detector in the
NO TE
user interface, see procedure below.
13 Check the firmware revision of the module. If it is older than the current
Entering the Serial Number using the Control Module
1 Connect the control module to the detector. Turn on the detector.
2 In the control module, press System (F5), then Records (F4). Using the
3 Press FW Update (F5). Now, press the m key. This will display a box which
4 Press Enter. This will display the box labeled Serial#.
5 Alphabetic characters and numbers are created using the up and down
6 Turn the detector off, then on again. The Records screen should display the
7 If a Agilent ChemStation is also connected, restart the Agilent ChemStation
1100 Series FD Reference Manual
Foams and Covers"
on page 196.
firmware revision, update the firmware using the standard firmware
update procedure, see
up/down arrows, make sure that the detector is highlighted.
says Update Enter Serial#.
arrows. In the box labeled Serial#, enter the 10-character serial number for
the detector. When the 10-character serial number is entered, press Enter
to highlight the complete serial number. Then, press Done (F6).
correct serial number for this module.
now as well.
Repairing the Fluorescence Detector
"Replacing the Detector's Firmware"
5
"Replacing the
on page 200.
161