5
Repairing the Fluorescence Detector
2 Use a 2.5 mm hex key to unscrew the 3 screws on the
assembly you want to remove.
4 Insert the new protection ring into its position. Do not
bend the protection ring.
188
3 Carefully remove the grating assembly and the protection
ring from its location.
5 For reinstallation route the cable first through the hole.
Carefully insert the grating into its position while pulling
the cable slightly from below. The couple spring has to be
clipped into the protection ring (TOP mark)!
1100 Series FD Reference Manual
Couple Spring
Protection ring