5
Repairing the Fluorescence Detector
10 Replace the lamp housing cover. Insert the optical unit
into its location and carefully insert the rear foam (A).
Then reconnect the cables.
Rear
foam
A
EM
A
A
B
EX
186
Next Steps:
•
Replace the covers as described in "Replacing the Foam
and the Top Cover" on page149.
•
Re-install the detector in the stack.
•
Reset the lamp counter as described in the user
interface documentation.
•
Turn the lamp on.
•
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
page 101.
B
"Wavelength Verification and Calibration"
1100 Series FD Reference Manual