Repairing the Fluorescence Detector

Replacing the Detector's Firmware

The installation of new firmware is required:
• if new version solves problems of currently installed version, or
• if after exchange of the main board (FLM) the version on board is older
To upgrade the detector's firmware the following steps have to be taken:
1 Load the firmware into the detector, see help system of your user interface.
2 If FLM board was replaced, reenter the serial number information of the
200
than the previously installed one.
module through the user interface, see
the Control Module"
Agilent ChemStation"
"Entering the Serial Number using
on page 161 or
"Entering the Serial Number using the
on page 162.