Advertisement

5
Repairing the Fluorescence Detector

Exchanging the Power Supply

When required
Tools required
Parts required
The repair of the power supply assembly comprises exchanging the complete assembly. No
NO TE
serviceable parts are inside.
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the cables.
Remove the detector from the stack and place it on the workbench.
Remove the front cover and top cover, see
Remove the processor board, see
Remove the fan assembly, see
Remove the optical unit, see
190
If defective
Screwdriver POZI 1 PT3
Wrench 1/4 inch
Wrench 5 mm
Wrench 7 mm
Power supply 0950-2528
"Removing the Covers"
"Exchanging the Processor Board"
"Exchanging the Fan"
on page 163.
"Removing the Optical Unit"
on page 144.
on page 159.
on page 178.
1100 Series FD Reference Manual