Firmware Update
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the FLD. Use the F5 key
(FW Update) to enter the Update section. If you want to update the resident
firmware (together with specific main firmware revisions), select the a file
from the PCMCIA card (RESnnnn.DLB) and press execute. If you want to
update the main firmware, press F7 key (Transfer) to turn the module into
the resident mode (LED on module should blink yellow).
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the FLD. Use the F5 key
(FW Update) to enter the Update section. Select the a file from the PCMCIA
card (1321nnnn.DLB) and press execute. When the update has finished the
update, press F7 key (Transfer) to return the module into the normal mode
(LED on module should stays yellow).
1100 Series FD Reference Manual
Introduction to the Control Module