Correcting Leaks - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector

Correcting Leaks

When required
Tools required
Parts required
1 Remove the front cover.
2 Use tissue to dry the leak sensor area and the leak pan.
3 Observe the capillary connections and the flow cell area for leaks and
4 Replace the front cover.
Figure 56
140
If a leakage has occurred in the flow cell area or at the capillary connections
Tissue
Two 1/4 inch wrenches for capillary connections
None
correct, if required.
Observing for Leaks
Leak sensor
1100 Series FD Reference Manual