Spectra Acquisition Modes - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Spectra Acquisition Modes

The spectra acquisition mode allows automatic storage of spectra during a
run. The mode can be changed during the run by time-programming.
The modes in detail are:
Table 57
1100 Series FD Reference Manual
Spectra Acquisition Modes
Mode
none
apex
all in peak
all
all without signals
Description
no spectra is stored
if apex of a peak is detected, one spectra is measured and stored.
In addition the next baseline spectra is stored.
The peak detectors will start and stop spectrum acquisition at begin
and end of a detected peak.
In addition the next baseline spectra is stored.
periodically all spectra are stored. The period depends on the
spectra range and step and the number of selected
chromatographical outputs (signals).
- for screening all flashes are used for spectra acquisition.
- no concentration connection is done.
- as chromatographical signal the mean of each spectra is
output/stored.
Theory of Operation
281
8