Exchanging the Leak Sensor or Leak Panel

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
Remove the flow cell.
1100 Series FD Reference Manual
If defective
Screwdriver POZI 1 PT3
Flat screwdriver
Leak sensor assembly 5061-3356
Leak Pan G1321-45501
1
"Removing the
Repairing the Fluorescence Detector
Locate the leak sensor cable.
5
165