Replacing the Lamp Cover Window

When required
Tools required
Parts required
When loosening or removing screws take care that they do not fall into the instrument. This
NO TE
may result in a complete removal of other assemblies.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell or disconnect the tubings from the
flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
1100 Series FD Reference Manual
If window is blocking the light throughput due to contamination
Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm long
Hexagonal screwdriver 4 mm, 100 mm long
Quartz Window 1000-0999
1
"Removing the
EM
EX
Repairing the Fluorescence Detector
Disconnect the emission and the excitation cables and
from the detector main board.
A
A
A
B
5
B
155