Replacing The Em And Ex Grating - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Replacing the EM and EX Grating

When required
Tools required
Parts required
When working on the optical unit, a clean workbench with ESD protection mat must be
CAU TI O N
available. Otherwise optical components or electronic boards may be damaged.
The grating assembly is shipped with a special transport packaging and includes a
NO TE
protection ring. DO NOT touch the grating surface and DO NOT bend the protection ring.
Preparations for this procedure:
Optical unit has been removed as described in
"Removing the Optical Unit"
The cables from the position encoders (at the rear of
the EM or EX grating assembly) should be
disconnected.
Disconnect the encoder cable of the
EX-monochromator from FLF board.
1100 Series FD Reference Manual
If defective
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
EM Grating assembly G1321-60004 (includes protection ring)
EX Grating assembly G1321-60003 (includes protection ring)
on page 178.
Repairing the Fluorescence Detector
1 Place the optical on the ESD protection mat and remove
the front foam (A). Locate the EM and EX grating
assembly.
EM
A
A
A
A
B
EX
5
B
187