Agilent Technologies 1100 Series Reference Manual page 175

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Remove the UV protector.
4 Use a pair of tweezers to replace the slit.
Next Steps:
Replace the UV protector.
Replace the front foam.
Replace the covers as described in
Re-install the flow cell and the front panel
Re-install the detector in the stack.
Turn the lamp on.
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
"Wavelength Verification and Calibration"
1100 Series FD Reference Manual
3 Locate the three slits (from left to right: PMT, EM, EX).
5 Put slight pressure on to the spring part of the slit.
"Replacing the Foams and Covers"
on page 101.
Repairing the Fluorescence Detector
on page 196.