Agilent Technologies 1100 Series Reference Manual page 180

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector
8 Replace the FLF board and fix the screws (A and B) and
connectors (C and D).
D
10 Reconnect the wires to the connector. Use a flat screw
driver to release the wire socket.
Next Steps:
Reinstall the optical unit as described in
If you replaced a G1321-66501 FLF board by a G1321-66511 FLF board, update the FLD firmware to revision A.03.7x or
above (to make it compatible).
182
A
B
C
"Installing the Optical Unit"
9 Check that the safety switch is positioned correctly. Look
through board and casting.
11 Put the wires into the released contacts using the
tweezers.
Order of wires (left to right):
green
empty
brown
yellow
white
on page 194.
1100 Series FD Reference Manual