Exchanging the Processor Board

When required
Tools required
Parts required
1 Turn off the lamp.
2 Switch off the module, and disconnect the cables.
3 Remove module from stack and place it on the workbench.
Analog
Analog
connector nuts
connector nuts
Remote
Remote
GPIB
GPIB
Board recesses
Board recesses
Figure 58
4 Remove the front cover, top cover and top foam section, see
5 Use a 5-mm and 7-mm wrench to unscrew the REMOTE and the GPIB
6 Disconnect all connectors from the processor board.
1100 Series FD Reference Manual
If detector main board is defective or for repair on other assemblies
Screwdriver POZI 1 PT3
Flat screw driver
Hexagonal wrenches 5 mm, 7 mm and 15 mm
Detector main board (FLM) G1321-69500 (exchange assembly)
Unscrew Connectors from Board
Covers"
on page 144.
connector and a 15-mm wrench to unscrew the nuts of the ANALOG
connectors.