2
Site Requirements and Specifications
Performance Specifications
Performance Specifications
Specifications
Performance Specifications G7117A
Table 2
Type
Detection type
Light source
Number of signals
Maximum data rate
Short term signal
noise (ASTM)
Drift
Linear absorbance
range
Wavelength range
Wavelength accuracy
Wavelength precision
Wavelength bunching
Slit width
Diode width
Time programmable
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Performance Specifications G7117A
Specification
1024-element photodiode array
Deuterium lamp
8
120 Hz (both spectra and signals)
With 10 mm Max-Light cartridge cell: <±3·10
4 nm, TC 2 s
With 60 mm Max-Light cartridge cell: <±0.6·10
4 nm, TC 2 s
-3
<0.5·10
AU/h at 230 nm
>2.0 AU (5 %) at 265 nm, typically 2.5 AU (5 %)
190 – 640 nm
±1 nm, self-calibration with deuterium lines
<±0.1 nm
2 – 400 nm, programmable in 1 nm steps
4 nm
~0.5 nm
Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength
range, threshold, spectra storage mode
-6
AU at 230 nm, slit width
-6
AU at 230 nm, slit width
30