Repairing the Fluorescence Detector
Note:
If an additional detector is added to the system, the fluorescence detector should be the last detector in the
flow path except for evaporative detectors, like LC-MSD. Otherwise the back pressure generated by the other
detector may overload the quartz flow cell and will lead to a defective cell (maximum pressure is 20 bar
(2 MPa).
Always use the outlet capillary set supplied with the accessory kit.
Note:
5
Replace the front cover.
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Note:
Perform a wavelength verification to check the correct positioning of the flow cell, as described in chapter
"Wavelength Verification and Calibration"
on page 101.
1100 Series FD Reference Manual