Rapidlab 1200 Operator's Guide: Troubleshooting
D70 Optics Error: 9
Problem
The system detects
a failure in the
CO-ox optical
components.
D70 Optics Error: 11, 12
Problem
The system detects
a failure in the
CO-ox optical
components.
D71–D77 Codes
D71 No Sample Detected at FD3
Problem
Fluid detector 3
(FD3) does not
detect the sample
during the
predefined time
limit.
CO-ox results are
not reported. Blood
gas, pH, electrolyte,
and metabolite
results are reported.
Corrective Action
1.
Select
Status > Shutdown
2.
Wait 10 seconds, and then turn the system on.
3.
If the problem persists, call for technical assistance.
Corrective Action
1.
Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
2.
Select
Diagnostics > CO-ox > Wavelength Cal
3.
Select
Status > Shutdown
4.
Wait 10 seconds, and then turn the system on.
5.
If the problem persists, call for technical assistance.
Corrective Action
•
Ensure the sample volume is sufficient for the sample type.
•
Select
Diagnostics > Fluidic Functions > RCx
•
Look for obstructions or poor flow in the CO-ox sample
path.
•
Remove obstructions from the CO-ox sample path:
If the obstruction is in the CO-ox sample tubing, replace the
tubing.
If the obstruction is in the CO-ox pump tubing or waste
tubing, remove the obstruction.
•
Repeat RCx test to check flow through the sample path.
•
Deproteinize the sample path.
Refer to Deproteinizing the Sample Path‚ page 5-4.
•
Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
•
If the problem persists, call for technical assistance.
.
.
.
.
02087462 Rev. V
6-71