Rapidlab 1200 Operator's Guide: Troubleshooting

D70 Optics Error: 9

Problem
The system detects
a failure in the
CO-ox optical
components.

D70 Optics Error: 11, 12

Problem
The system detects
a failure in the
CO-ox optical
components.
D71–D77 Codes

D71 No Sample Detected at FD3

Problem
Fluid detector 3
(FD3) does not
detect the sample
during the
predefined time
limit.
CO-ox results are
not reported. Blood
gas, pH, electrolyte,
and metabolite
results are reported.
Corrective Action
1.
Select
Status > Shutdown
2.
Wait 10 seconds, and then turn the system on.
3.
If the problem persists, call for technical assistance.
Corrective Action
1.
Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
2.
Select
Diagnostics > CO-ox > Wavelength Cal
3.
Select
Status > Shutdown
4.
Wait 10 seconds, and then turn the system on.
5.
If the problem persists, call for technical assistance.
Corrective Action
Ensure the sample volume is sufficient for the sample type.
Select
Diagnostics > Fluidic Functions > RCx
Look for obstructions or poor flow in the CO-ox sample
path.
Remove obstructions from the CO-ox sample path:
If the obstruction is in the CO-ox sample tubing, replace the
tubing.
If the obstruction is in the CO-ox pump tubing or waste
tubing, remove the obstruction.
Repeat RCx test to check flow through the sample path.
Deproteinize the sample path.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
If the problem persists, call for technical assistance.
.
.
.
.