Siemens Rapidlab 1200 Operator's Manual page 399

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
Table E-37
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
*
WRSD = within-run standard deviation
Rapidlab 1260 Recovery and Precision Testing—Na
N
WRSD
Capillary
24
0.26
Capillary
24
0.21
Capillary
24
0.30
Capillary
24
0.55
Capillary
24
0.43
Capillary
24
0.85
Syringe
47
0.27
Syringe
48
0.29
Syringe
48
0.34
Syringe
48
0.71
Syringe
48
0.34
Syringe
48
0.78
Micro-
12
0.62
capillary
Micro-
12
0.18
capillary
Micro-
12
0.60
capillary
Micro-
12
1.11
capillary
Micro-
12
0.29
capillary
Micro-
12
1.29
capillary
Micro-
24
0.41
syringe
Micro-
24
0.15
syringe
Micro-
23
0.21
syringe
Micro-
24
0.62
syringe
Micro-
24
0.47
syringe
Micro-
22
0.99
syringe
*
Observed
Expected
113.8
111.2
130.6
129.4
135.0
134.4
142.8
141.5
147.9
148.6
175.7
179.8
114.3
111.2
129.7
129.4
134.8
134.4
141.2
141.5
148.9
148.6
174.6
179.8
110.2
109.9
129.5
130.1
131.7
132.0
143.1
143.8
149.6
150.8
175.4
178.1
109.6
109.9
130.6
130.1
132.1
132.0
143.5
143.8
150.1
150.8
178.7
178.1
E-39
+
Recovery
WRCV
102.3
0.2
100.9
0.2
100.4
0.2
100.9
0.4
99.5
0.3
97.7
0.5
102.7
0.2
100.2
0.2
100.3
0.3
99.8
0.5
100.2
0.2
97.1
0.4
100.3
0.6
99.6
0.1
99.8
0.5
99.5
0.8
99.2
0.2
98.5
0.7
99.8
0.4
100.4
0.1
100.1
0.2
99.8
0.4
99.5
0.3
100.3
0.6