5-18
d. Squeeze the finger snaps on the reagent manifold and then remove the manifold
1
Figure 5-8
e. Dispose of the manifold according to your institution's protocol.
4. Install the new reagent manifold:
a. Snap the new reagent manifold in place.
b. Connect the waste tubing.
c. For Rapidlab 1245 and 1265 systems, connect the reagent manifold tubing to the
Completing Reagent Manifold Maintenance
1. Close the front cover.
2. Reinstall the reagent and wash cartridges.
3. Select
4. Verify system performance:
a. Perform a 2-point calibration.
b. Analyze a minimum of 2 levels of quality control material.
02087462 Rev. V
by sliding it down and away from the sample connector.
Finger snaps
Reagent Manifold Finger Snaps
When you are installing the tubing, avoid pinching or bending the tubing.
NOTE:
CO-ox pump tubing at the fitting.
.
Return
Refer to Performing Manual Calibrations‚ page 3-4.
Rapidlab 1200 Operator's Guide: Maintenance