Performing Reference Sensor Maintenance

The Reference sensor requires regular cleaning to maintain performance.

Cleaning and Inspecting the Reference Sensor

You need the following materials for this task:
Water
Lint-free swab
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Remove the Reference sensor.
Refer to Removing the Sensors‚ page 5-30.
1
2
Figure 5-16
02087462 Rev. V
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module
Reference sensor
Spring-loaded latch
Measurement Module
Rapidlab 1200 Operator's Guide: Maintenance