Flushing The Tubing For The Automaticqc Module - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Maintenance

Flushing the Tubing for the AutomaticQC Module

1. Connect the syringe tubing to the AutomaticQC manifold.
2. Flush with a minimum of 3 mL water.
1
2
Figure 5-31
3. Flush with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
5. Disconnect the syringe tubing.
Water collection area
AutomaticQC manifold
AutomaticQC Tubing
5-57
02087462 Rev. V