Replacing boards and components

Replacing the Compact optics

Fig. 32:
WARNING
WARNING
CAUTION
© Siemens, 2008
For internal use only
Electrical voltage!
See
(Safety information / p.
Switch the SIREMOBIL Compact off and wait approxi-
mately 3 minutes for the voltage in the I.I. mini-voltage
supply to drop.
Risk of injury on mechanical parts!
If these guidelines are not followed, minor to moderate injury,
especially to the hands, can occur.
Follow the notes concerning the C-arm carbon fiber
structure.
Risk of injury!
Noncompliance can lead to minor to medium injury and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg. Hold the covers firmly when installing or remov-
ing them.
Before the C-arm cover is loosened, the C-arm must be
moved to its middle orbital position. If the C-arm cov-
ers are loosened, the C-arm may not be moved to its
orbital end positions (I.I. or SIREPHOS at the orbital
end position).