02087462 Rev.V
Rapidlab 1260 Recovery and Precision Testing—pCO
N
WRSD
Capillary
24
0.50
Capillary
24
0.34
Capillary
24
0.45
Capillary
24
0.83
Capillary
24
0.90
Capillary
24
1.16
Capillary
24
1.53
Capillary
24
3.41
Syringe
45
0.28
Syringe
48
0.35
Syringe
48
0.37
Syringe
48
0.81
Syringe
45
0.60
Syringe
48
0.86
Syringe
48
1.27
Syringe
45
2.16
Syringe
47
5.09
Micro-
12
0.25
capillary
Micro-
12
0.28
capillary
Micro-
12
0.28
capillary
Micro-
12
1.03
capillary
Micro-
12
0.62
capillary
Micro-
12
1.24
capillary
Micro-
12
1.66
capillary
Micro-
12
1.71
capillary
Micro-
12
4.66
capillary
Micro-
24
0.29
syringe
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
22.6
21.3
29.0
28.4
36.9
35.4
48.3
49.9
72.0
71.1
99.3
99.1
129.9
135.4
216.3
226.4
7.5
7.1
21.6
21.4
28.9
28.6
36.6
35.5
49.6
49.9
72.7
71.4
95.5
99.3
130.6
135.4
213.4
227.0
6.9
7.1
21.0
21.3
27.9
28.4
35.3
35.8
48.5
49.8
70.5
71.1
100.2
100.2
137.7
135.1
230.2
229.1
6.7
7.1
2
Recovery
WRCV
105.9
2.2
101.9
1.2
104.2
1.2
96.8
1.7
101.3
1.3
100.2
1.2
95.9
1.2
95.6
1.6
105.8
3.7
100.6
1.6
101.2
1.3
103.1
2.2
99.4
1.2
101.7
1.2
96.2
1.3
96.5
1.7
94.0
2.4
96.9
3.7
98.4
1.3
98.2
1.0
98.7
2.9
97.4
1.3
99.2
1.8
100.0
1.7
101.9
1.2
100.5
2.0
94.6
4.3