Conditioning The Sensors - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

5-6

Conditioning the Sensors

Conditioning cleans and conditions the glass membranes of the pH and sodium sensors.
You need the following materials for this task:
Conditioner
Syringe
Glucose and Lactate test/blank sensors (TB4)
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
conditioner, or deproteinizer. Replace the biosensors with the test/blank sensors
(TB4) before conditioning the sensors or deproteinizing the sample path. Reinstall
the biosensors within 2 hours.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. At user interface, select
2. Select
3. Select
4. Select the video button and follow the instructions in the video.
5. If you have a Rapidlab 1260 or 1265 system, replace the Glucose and Lactate
biosensors with the test/blank sensors:
a. Select the video button and follow the instructions in the video.
b. Select
02087462 Rev. V
Wear personal protective equipment. Use universal precautions.
Do not expose the glucose and lactate biosensors to bleach,
Do not remove or return the sensors to the measurement module
Status > Maintenance
Conditioning the Sensors
.
Continue
.
Continue
Rapidlab 1200 Operator's Guide: Maintenance
.
.