Rapidlab 1200 Operator's Guide: Maintenance
10. Start up the system:
a. Turn the power switch on.
b. At the interface screen, use the numeric buttons to enter your password.
c. Select
Replacing the CO-ox Roller Cage
Use this procedure to replace the roller cage of the CO-ox pump.
You need the following materials for this task:
•
Lint-free tissue and swabs
•
Roller cage
•
Water
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
Removing the Roller Cage
1. Select
2. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
3. Disconnect the reagent manifold tubing from the CO-ox pump tubing at the
fitting.
1
2
Figure 5-22
.
Continue
Wear personal protective equipment. Use universal precautions.
Status > Maintenance
Fitting
Reagent manifold tubing
CO-ox Module and Reagent Manifold
.
5-45
02087462 Rev. V