Replacing The Co-Ox Lamp - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Maintenance

Replacing the CO-ox Lamp

WARNING:
Waiting for at least 5 minutes allows sufficient time for the lamp to cool.
You need a CO-ox lamp for this task.
1. Ensure that the system is not performing any analysis.
CAUTION:
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
WARNING:
this procedure to shut down the system.
2. Shutdown the system:
a. Select
b. When prompted, turn the power switch off.
3. Move the system so that you are able to work directly behind the system.
4. For ease of access, prop the back of the system up by about 3/4 inch (1.9 cm).
When you prop up the back of the system, the back cover can lay flat when
you open it.
Do not touch the lamp until it has been off for at least 5 minutes.
Do not remove power from the system for more than 6 hours if
To prevent damage to the hard drive and permanent loss of data, use
Status > Shutdown > Yes
The power switch is on the back panel of the system.
.
5-43
02087462 Rev. V