Rapidlab 1200 Cartridges - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

C-2

Rapidlab 1200 Cartridges

Each of the cartridges has an install-by date on the label.
Cartridge
Reagent
cartridge
Wash Cartridge
Rapidlab 1245
and 1265
Wash Cartridge
Rapidlab 1240
and 1260
AutomaticQC
cartridge
02087462 Rev. V
Rapidlab 1200 Operator's Guide: System Fluids
Stability
Use life
30 days
Calibration dependent
18 days
Sample dependent
30 samples per day = 12
days of use
18 days
Sample dependent
30 samples per day = 17
days of use
28 days
sample dependent
at least 1 sample analysis
of each level, 3 times per
day
Storage temperature
°
2
to 8°C
Room temperature not to
exceed 25°C, up to 7 days
Room temperature at
approximately 25°C
Room temperature at
approximately 25°C
°
2
to 8°C