Siemens Rapidlab 1200 Operator's Manual page 368

Table of Contents

Advertisement

E-8
Analyte (units)
*
Glucose system performance will be out of specification when PAM levels exceed
16 µg/mL.
Lactate system performance will be out of specification when PAM levels exceed
64 µg/mL.
02087462 Rev.V
Rapidlab 1200 Operator's Guide: System Specifications
Interference
Interference
Level Tested
Sodium
70 mg/dL
Nitroprusside
Thiocyanate
80 mg/dL
Heparin
20,000 U/dL
Epinephrine
2 mg/dL
Norepinephrine
2 mg/dL
Phenobarbital
15 mg/dL
Glutamate
16 mg/dL
Hetastarch
30%
Acetoacetate
40 mg/dL
Ascorbate
8 mg/dL
Dilantin
14 mg/dL
Bilirubin (Direct)
30 mg/dL
Bilirubin (Total)
35 mg/dL
Pralidoxime Iodide
64 µg/mL
Creatinine
30 mg/dL
Hydroxybutyrate
200 mg/dL
Urea
500 mg/dL
Guaiacol
5 mg/dL
Pyruvate
9 mg/dL
Theophylline
9 mg/dL
Penicillamine
25 mg/dL
Isoniazid
2 mg/dL
Uric Acid
10 mg/dL
Citrate
1000 mg/dL
Potassium Oxalate
1000 mg/dL
EDTA
800 mg/dL
Sodium Fluoride
1000 mg/dL
Sodium Fluoride/
1000 mg/dL
Potassium Oxalate
Acetominophen
2 mg/dL
Effect of
Interference
(Test - control)
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
< +/-0.3
1.0
1.0
0.35

Advertisement