Replacing the Measurement Module Tubing

You need a tubing kit for this task.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1. Select
2. Select
3. Lift the door on the measurement module.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
4. Replace the pinch valve tubing.
Perform steps 4 and 5 of Replacing the Pinch Valve Tubing‚ page 5-10.
5. Replace the waste tubing between the pressure sensor bubbler and the reagent
manifold:
a. Disconnect the waste tubing from the pressure sensor bubbler and the reagent
1
2
3
Figure 5-3
02087462 Rev. V
Wear personal protective equipment. Use universal precautions.
Status > Diagnostics > Cartridges
.
Eject R Cartridge
manifold.
Pressure sensor bubbler
Waste tubing
Reagent manifold
Measurement Module–Rapidlab 1260 and 1265 systems
Rapidlab 1200 Operator's Guide: Maintenance
.