Siemens Rapidlab 1200 Operator's Manual page 404

Table of Contents

Advertisement

E-44
Table E-41
Level Mode
4
5
1
2
3
4
5
*
WRSD = within-run standard deviation
Table E-42
Level Mode
1
2
3
4
5
1
2
3
4
5
1
2
02087462 Rev.V
Rapidlab 1260 Recovery and Precision Testing—Glucose
N
WRSD
pH, Glu,
24
4.75
Lac
Capillary
pH, Glu,
23
14.82
Lac
Capillary
pH, Glu,
48
1.18
Lac
Syringe
pH, Glu,
48
2.01
Lac
Syringe
pH, Glu,
43
2.91
Lac
Syringe
pH, Glu,
48
4.93
Lac
Syringe
pH, Glu,
48
6.15
Lac
Syringe
Rapidlab 1260 Recovery and Precision Testing—Lactate
N
WRSD
Capillary
21
0.09
Capillary
24
0.06
Capillary
21
0.08
Capillary
24
0.16
Capillary
21
0.17
Syringe
45
0.03
Syringe
48
0.05
Syringe
45
0.15
Syringe
48
0.20
Syringe
45
0.21
Micro-
12
0.09
capillary
Micro-
9
0.09
capillary
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
353
365
500
525
48
43
93
87
216
213
377
382
500
520
*
Observed
Expected
1.5
1.5
2.1
2.2
4.7
5.1
9.1
11.0
11.7
14.2
1.1
1.3
2.1
2.2
5.1
5.6
9.3
10.9
12.1
14.3
1.1
1.0
4.3
4.4
Recovery
WRCV
97.0
1.3
95.2
3.0
112.8
2.5
107.0
2.2
101.6
1.3
98.6
1.3
96.1
1.2
Recovery
WRCV
97.5
6.2
96.8
3.0
91.4
1.7
83.2
1.8
82.4
1.5
89.8
3.0
94.6
2.4
90.5
3.1
85.4
2.2
84.3
1.7
104.4
8.1
98.4
2.2

Advertisement

Table of Contents
loading

Table of Contents