D8 Po R Cartridge Error; D14 No Sample Detected At Fd2; D19 Fluid Detector Error: 1, 2, 3; D8–D22 Codes - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

6-58
D8–D22 Codes
D8 pO
Problem
The pO2 value
during a calibration
is different than the
assigned value.

D14 No Sample Detected at FD2

Problem
Fluid detector 2 (FD2) does
not detect the sample during
the predefined time limit.
The sample is discarded and
no results are reported.

D19 Fluid Detector Error: 1, 2, 3

Problem
Fluid detector millivolt
(mV) reading is beyond
predefined limits.
1 FD1
2 FD2
3 FD3
02087462 Rev. V
R Cartridge Error
2
Corrective Action
Calibrate the Barometric Sensor. Refer to Calibrating the
Barometric Sensor‚ page 5-3
Perform a full calibration. Refer to Performing Manual
Calibrations‚ page 3-4.
If the pO
sensor.
Perform a full calibration. Refer to Performing Manual
Calibrations‚ page 3-4.
If the problem persists, replace the reagent cartridge.
Perform a full calibration. Refer to Performing Manual
Calibrations‚ page 3-4.
If the problem persists, call for technical assistance.
Corrective Action
Corrective Action
Rapidlab 1200 Operator's Guide: Troubleshooting
sensor is greater than 6 months old, replace the
2
Perform a full calibration.
Refer to Performing Manual Calibrations‚
page 3-4.
Repeat sample analysis.
If the problem persists, call for technical assistance.
Perform a full calibration.
Refer to Performing Manual Calibrations‚
page 3-4.
If the problem persists, call for technical assistance.