Rapidlab 1200 Operator's Guide: Maintenance

Replacing the Reagent Manifold

You need a reagent manifold for this task.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1. Select
2. Select
3. Remove the reagent manifold:
a. Lift the front cover.
b. Disconnect the waste tubing from the reagent manifold.
c. For Rapidlab 1245 and 1265 systems, disconnect the reagent manifold
1
2
3
Figure 5-7
Wear personal protective equipment. Use universal precautions.
Status > Diagnostics > Cartridges
Eject Both Cartridges
Refer to Preparing for Maintenance Procedures‚ page 5-1.
tubing from the CO-ox pump tubing at the fitting.
Do not disconnect the reagent manifold tubing from the reagent manifold.
Fitting
Reagent manifold tubing (Rapidlab 1245 and 1265 systems)
Waste tubing
Sample Chamber and Reagent Manifold
.
, remove the cartridges, and set them aside.
5-17
02087462 Rev. V