Rapidlab 1200 Operator's Guide: Troubleshooting
b. Reinsert the reagent manifold tubing (item 2, Figure 6-7) into the fitting
c. Reinsert the waste tubing into the port on the top, right-side of the reagent
9. Reinstall the pO
a. Push the black spring enclosure assembly to the right, until the
b. Remove the syringe.
c. Remove the clot clearing tool, by lifting the clot clearing tool lip up at the
d. Reinstall the pO
e. Press the black spring enclosure assembly lever to reapply pressure to the
10. Close the system.
a. Reinstall the reagent and wash cartridges.
b. Close the reagent and wash module doors.
c. Close the front cover of the system.
11. Select the Return button twice to exit Diagnostics mode.
You are prompted to calibrate the system.
12. Select
The clot clearing procedure is complete.
(item 1, Figure 6-7) that connects to the CO-ox pump tubing
(Rapidlab 1245 and 1265 systems).
manifold (item 4, Figure 6-7).
If an obstruction has been cleared, go to step 9.
If an obstruction has not been cleared, go to
the Reagent Cartridge‚ page 6-22
sensor.
2
mechanism clicks in place.
bottom while pressing the top of the tool inward.
sensor.
2
sensor assembly.
to calibrate the system.
Yes
Clearing the Clot from
.
6-21
02087462 Rev. V