Rapidlab 1200 Operator's Guide: Maintenance
Replacing the AutomaticQC Manifold
You need an AutomaticQC manifold for this task.
1. Press down on the left side of the AutomaticQC cartridge connector and slide
the connector to the right.
2. Select
3. Select
4. Select
5. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
6. Remove the AutomaticQC manifold:
a. Pull the connector port cover towards you.
1
2
3
Figure 5-9
b. Push down on the pins on the manifold and pull the manifold and
7. Install the new AutomaticQC manifold:
a. Insert the bottom clip into the opening on the system.
.
Continue
Diagnostics > Cartridges
, remove the reagent cartridge, and set it aside.
Eject R Cartridge
Connector port cover
Cartridge connector
Pins
Connector Port and Cover
connector port away.
.
5-19