Rapidlab 1200 Operator's Guide: Maintenance

Cleaning and Drying the Tubing

Ensure that you follow these instructions using at least the minimum volumes of
fluid and air described in these procedures.

Flushing the Tubing for the RCx Reagent

1. Connect a piece of tubing to a syringe.
2. Connect the syringe tubing to the connector for the RCx reagent.
1
2
Figure 5-25
3. Flush with a minimum of 3 mL water.
4. Flush with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
Water collection area
RCx reagent Connector
Reagent Manifold
5-51
02087462 Rev. V