D23 Reagent Flow Error: 14; D24 Codes; D24 Aqc Material Error: 1, 2, 3; D32 Codes - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting

D23 Reagent Flow Error: 14

Problem
The system detects
a possible
obstruction during
calibration or
during the wash
cycle after sample
analysis.

D24 Codes

D24 AQC Material Error: 1, 2, 3

Problem
An error was
detected during
delivery of AQC
material.
1 level 1
2 level 2
3 level 3

D32 Codes

D32 AQC Cartridge Valve Error: 1

Problem
A valve does not
move to the correct
position.
Corrective Action
1.
Select
Diagnostics > Fluidic Functions > RCx
obstructions.
2.
Remove obstructions from reagent manifold, measurement
module, and sample tubing.
3.
Resolve any other D codes.
4.
Perform the RCx test again.
5.
Select
Diagnostics > Fluidic Functions > Wash Cycle
6.
If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
.
Functions
7.
If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
.
Functions
8.
If the problem persists, call for technical assistance.
Corrective Action
1.
Check that the AutomaticQC connector is completely closed
and that the sensors are aligned correctly.
2.
Select
Diagnostics > Fluidic Functions > Prime AQC
.
Cartridge
The correct button is highlighted.
3.
Check if there is an obstruction in the reagent probe
cartridge.
Refer to Flushing the Tubing for the AutomaticQC Module‚
page 5-57.
4.
If the problem persists, replace the AutomaticQC cartridge.
If you are not able to replace the cartridge, turn off the
NOTE:
AutomaticQC option in Setup to continue analyzing patient
samples.
5.
If the problem persists, call for technical assistance.
Corrective Action
1.
Select
Diagnostics > Pumps and Valves > AQC Valve
2.
If the problem persists, replace the AutomaticQC cartridge.
3.
If the problem persists, call for technical assistance.
6-63