Siemens Rapidlab 1200 Operator's Manual page 423

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
Table E-61
Level Mode
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
Rapidlab 1265 Recovery and Precision Testing—Glucose
N
WRSD
Syringe
40
1.57
Syringe
45
2.02
Syringe
45
2.24
Syringe
42
5.25
Micro-
12
2.02
capillary
Micro-
12
2.47
capillary
Micro-
12
1.41
capillary
Micro-
12
7.42
capillary
Micro-
12
4.21
capillary
Micro-
24
1.76
syringe
Micro-
24
2.05
syringe
Micro-
24
1.87
syringe
Micro-
24
18.80
syringe
Micro-
24
8.40
syringe
pH, Glu,
21
0.83
Lac
Capillary
pH, Glu,
24
1.80
Lac
Capillary
pH, Glu,
24
2.60
Lac
Capillary
pH, Glu,
24
3.93
Lac
Capillary
pH, Glu,
21
9.84
Lac
Capillary
*
Observed
Expected
87
82
214
211
339
358
477
472
43
44
86
88
224
219
332
343
447
452
48
48
99
98
226
222
367
374
456
452
52
44
98
93
203
205
355
365
495
526
E-63
Recovery
WRCV
105.5
1.8
101.2
0.9
94.9
0.7
101.1
1.1
96.8
4.7
97.9
2.9
102.4
0.6
96.7
2.2
98.8
0.9
101.0
3.7
101.1
2.1
102.1
0.8
98.1
5.1
100.9
1.8
116.5
1.6
106.3
1.8
99.4
1.3
97.3
1.1
94.2
2.0
02087462 Rev.V