D32 AQC Cartridge Valve Error: 2

Problem
The AutomaticQC
cartridge fails to
eject.

D33 Codes

D33 R Cartridge Valve Error

Problem
The valve in the
reagent cartridge
fails to move to the
correct position.

D34 Codes

D34 Waste Detector Cal Error: 1

Problem
The waste detector
is out of range
during a system
calibration.

D34 Waste Detector Cal Error: 2, 3

Problem
A full or empty
waste bottle was
inserted at the
wrong time during
waste bottle
calibration.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
Corrective Action
1.
Check that the connector is completely open and then select
to eject the cartridge.
Replace
2.
Select
Diagnostics > Cartridges > Eject AQC Cartridge
If you are not able to replace the cartridge, in Setup,
NOTE:
turn the AutomaticQC option off to continue analyzing patient
samples.
3.
If the problem persists, call for technical assistance.
Corrective Action
1.
Select
Diagnostics > Pumps and Valves > R Valve
2.
Select
Diagnostics > Cartridges > Eject Reagent
.
Cartridge
3.
If the problem persists, replace the reagent cartridge.
4.
If the problem persists, call for technical assistance.
Corrective Action
1.
Select
Diagnostics > Waste Detector
2.
Select
Status > Shutdown
3.
Wait 10 seconds, and then turn the system on.
4.
If the problem persists, call for technical assistance.
Corrective Action
Select
Diagnostics > Waste Detector
Be sure to insert the waste bottle either full or empty at the
appropriate times during the calibration.
If the problem persists, call for technical assistance.
.
.
to shut down the system.
.
.